Show simple item record

dc.contributor.authorRaj, P.en_US
dc.contributor.authorBall, J.L.en_US
dc.contributor.authorCarmichael, J.en_US
dc.contributor.authorFrenje, Johan A.en_US
dc.contributor.authorGocht, R.en_US
dc.contributor.authorGorini, G.en_US
dc.contributor.authorHolmes, I.en_US
dc.contributor.authorGatu Johnson, Mariaen_US
dc.contributor.authorKennedy, R.en_US
dc.contributor.authorMackie, S.en_US
dc.contributor.authorNoncente, M.en_US
dc.contributor.authorPanontin, E.en_US
dc.contributor.authorPetruzzo, M.en_US
dc.contributor.authorRebai, M.en_US
dc.contributor.authorReinke, M.en_US
dc.contributor.authorRice, John E.en_US
dc.contributor.authorRigamonti, D.en_US
dc.contributor.authorDalla Rosa, M.en_US
dc.contributor.authorSaltos, A.A.en_US
dc.contributor.authorTardocchi, M.en_US
dc.contributor.authorTinguely, R. Alexen_US
dc.contributor.authorWang, X.en_US
dc.date.accessioned2025-03-21T20:22:12Z
dc.date.available2025-03-21T20:22:12Z
dc.date.issued2024-09
dc.identifier24ja075
dc.identifier.urihttps://hdl.handle.net/1721.1/158720
dc.descriptionSubmitted for publication in Review of Scientific Instruments
dc.description.abstractNeutron measurement is the primary tool in the SPARC tokamak for fusion power (Pfus) monitoring, research on the physics of burning plasmas, validation of the neutronics simulation workflows, and providing feedback for machine protection. A demanding target uncertainty (10% for Pfus) and coverage of a wide dynamic range (>8 orders of magnitude going up to 5x10^19 n/s), coupled with a fast-track timeline for design and deployment, make the development of the SPARC neutron diagnostics challenging. Four subsystems are under design, which exploit the high flux of direct DT and DD plasma neutrons emanating from a shielded opening in a midplane diagnostic port. The systems comprise: a set of ~15 flux monitors mainly ionization chamber and proportional counters for measurement of the neutron yield rate, two independent foil activation systems for measurement of the neutron fluence, a spectrometric radial neutron camera for poloidal profiling of the plasma emissivity, and a high-resolution magnetic proton recoil spectrometer for measurement of the core neutron spectrum. Together, the four systems ensure redundancy of sensors and methods, and aim to provide high resolutions of time (10 ms), space (~7 cm), and energy (<2% at 14 MeV). This paper presents the broader objectives behind the preliminary design of the SPARC neutron diagnostics, and discusses the ongoing studies on neutronics, detector comparisons, prototyping, and integration with the unique infrastructure of SPARC. Engineering details of the four subsystems and the concepts for in-situ neutron calibration are also highlighted.
dc.publisherAIPen_US
dc.relation.isversionofdoi.org/10.1063/5.0219538
dc.sourcePlasma Science and Fusion Centeren_US
dc.titleOverview of the Neutron Diagnostic Systems for the SPARC Tokamaken_US
dc.typeArticleen_US
dc.contributor.departmentMassachusetts Institute of Technology. Plasma Science and Fusion Center
dc.relation.journalReview of Scientific Instruments